Product Details:
|
Package: | 24 Tests/kit | Expiration Date: | 12 Months |
---|---|---|---|
Storage: | -20±5℃ | Applicable: | Micgene 244/244 IVD(BIO-KEY) |
LoD: | 5 Copies/μL | CV: | ≤5% |
High Light: | Real Time Herpes Simplex Virus PCR,24tests/Kit Herpes Simplex Virus PCR,HSV-6 PCR Detection Kit |
Herpes simplex virus 6 (HSV-6) real-time PCR detection kit (lyophilized)
--24 tests/kit
Intended use:
This kit, adopting Real-Time PCR-Fluorescence technology, is used for detection of herpes simplex virus 6 (HSV-6). It can be used for the laboratory diagnosis and monitoring of the herpes simplex virus 6 (HSV-6) infection. The test results are only for clinical reference, not for case confirmation or exclusion.
Principle:
The specific primers and specific fluorescent probes are designed for the nucleic acid conserved region of herpes simplex virus 6 (HSV-6). The PCR reaction solution and real-time fluorescent quantitative PCR detection technology is applied on the fluorescence quantitative PCR instrument to realize the rapid detection of herpes simplex virus 6 (HSV-6) through the change of fluorescence signal. The detection system contains a human internal control (IC) gene, which is used to monitor whether the specimens and the process of nucleic acid extraction is qualified.
Main components:
No. | Components |
BIK-QL-H0014S1 |
|
PCR Enzyme Mix (Lyophilized) |
24 tests/bottle |
2 | Positive Control | 100 μL/tube |
3 | Negative Control | 100 μL/tube |
4 | RNase-free water | 1mL/tube |
5 | Paraffin oil | 1.5mL/tube |
Note: Do not mix the components from different batches for detection. The Positive Control of HSV-6 and internal control were constructed artificially, and they were not infectious.
Product lyophilization package:
Specimen requirements:
1.Specimen type :
Urethral swabs, cervical swabs or rectal swab.
2.Specimen preservation:
The specimen can be stored for 3 months at -20±5℃ and for a long time below -70℃.
Reaction mix preparation:
Refer to the tables below to prepare the reaction mix:
1×volume required | |
Resuspended master mix | 14 μL |
HSV-6 Primer & probe |
1 μL |
Total volume | 15 μL |
※Multiply the numbers according to the number of tests.
Quality control:
Negative Control, Positive Control and IC of samples must be performed correctly, otherwise the sample results is invalid.
Ct value | |
Negative Control | No Ct |
Positive Control | ≤30 |
IC of samples | ≤38 |
Operation details please check from IFU!